1. Reflector system designed in CAD; color temperature close to that of sunlight; making the tissue surface visible under natural color; depth of tubular light beams up to over 600mm; uniform illumination effectively eliminating shadows under any circumstance;
2. Special heat-isolated radiation filter reduces heat to the minimum level.
3. Rippled panel made of shock-resistant and heat-resistant PC material, making light soft, uniform and increasing safety.
4. Handle type CPU control. 8 stages light adjustment.
5. The major and auxiliary lamps can switch automatically.
6. The major and auxiliary bulbs can be replaced easily. The dischargeable handle sheath can be sterilized in steam sterilizer tungsten halogen lamps of international brand are adopted with the average service life up to 1000 hours.
7. Each balance arm has two 360 degree rotatable joints.
Specifications:
Illumination lx(at the distance of 1mm) > 120000> 120000
Color Temperature 4300±500K
Diameter of Spot 150~300mm
Depth of light bean ≥600mm
Brightness adjustment 8-stage continuous light adjustment
Temperature rise on the head of patient under surgical operation ≤2°C
Diameter of lamp holder 700/700mm
Shift of major and standby lamps Automatic
Bulb voltage 24V
Bulb power 150-700 W /150-700 W
Power supply 220±10%V, 50-60Hz
Input power 460W
Minimum height for installation mm 2900